Data Sheet - Acid Digestion Vessels

Part. No. CapacityBody
Dia
Body
Height
Closure
Type
Cups
Inner
Dia
Cup
Inner
Height
Body
Moc
LinerMax
PSI
Max
Temp.
75645 10 ml 65 mm 83 mm Spanner Wrench 26 mm 23 mm SS 316Teflon1800 240°C
7564625 ml 65 mm 83 mm Spanner Wrench 26 mm 44 mm SS 316Teflon1800 240°C
75647 25 ml 60 mm 86 mm Six Screws Cap 31 mm 30 mm SS 316Teflon1200240°C
7564825 ml 84 mm 126 mm Six Screws Cap 31 mm 32 mm SS 316Teflon5000240°C
7564925 ml 80 mm 112 mm Six Screws Cap 31 mm 32 mm SS 316PPL 1200 280°C
7565050 ml 64 mm 134 mm Spanner Wrench 26 mm 95 mm SS 316Teflon1800240°C
7565150 ml 80 mm 143 mm Six Screws Cap 31 mm 62 mm SS 316PPL 1200 280°C
75652125 ml95 mm 160 mm Six Screws Cap 46 mm 83 mm SS 316Teflon1900 240°C
75653200 ml 95 mm 210 mm Six Screws Cap 46 mm 135 mm SS 316Teflon1900 240°C

Heating Blocks Optional | Available with PPL liner on Special Request | PPL ( polyphenylene polymer) Liner Max Working Temp 280°C

For Over Pressure : A safety blow-off disc above the PTFE cup will release pressure

We produce both standard and customized autoclaves with variations in width, depth, height and volume.

Overview

The Acid Digestion Vessel is specifically designed for the preparation of laboratory samples for chemical analysis. It enables the digestion or dissolution of organic and inorganic samples in strong acids or alkalis under elevated temperatures and pressures. Available in both general-purpose and microwave-compatible variants, these vessels ensure complete containment and recovery of samples, making them ideal for accurate analytical testing.

How It Works

To perform sample digestion, place the sample inside the vessel and add the required acid or alkali. Seal the vessel securely to maintain pressure containment. Heat the vessel to the desired temperature, allowing the chemical reaction to fully dissolve the sample. Once digestion is complete, carefully release the pressure and recover the solution for analysis.

These vessels are essential for laboratories conducting precise elemental analysis, offering enhanced safety, efficiency, and reliability in sample preparation. 

For further details, feel free to